Eppendorf Reference® 2 可调量程移液器
Reference代表着非凡的精确性和准确性、经久耐用及人体工程学设计。新品 Reference 2 移液器以其出色的产品特性和创新的技术,必将成为您工作上可信赖的伙伴。
- 产品特性
- 单键操作
人体工程学设计,操作用力小
吸头脱卸前有触觉反馈
吸头脱卸简单快捷,减少气溶胶
不必移动拇指,减少重复性劳损
特殊表面处理边缘光滑
握感舒适
清洁简便
一体化的部件使用寿命长,经久耐用
序列号多重提示位于手柄和移液器下半部
防止体积显示部件装配错误
配有弹性吸嘴提高移液均一性
降低操作/脱卸用力
移液器特点操作按钮具有彩色标识和量程标签,可迅速识别量程和匹配吸头
上半部圆形设计,无手指凹槽,可任何角度操作
精确的量程设定:四位数字体积显示(各个角度清晰可见)
体积设置快速安全,带量程锁
可整支高温高压灭菌
边缘不锈钢环保护,极其坚固
产品技术参数:
产品型号 | 体积 | 不准确度 | 不精确度 | |||
± % | ± μL / mL | ± % | ± μL / mL | |||
Eppendorf Reference® 2,单道移液器,可调量程* | ||||||
| 0.1 μL | ±48.0 | ±0.048 | ±12.0 | ±0.012 | |
0.25 μL | ±12.0 | ±0.03 | ±6.0 | ±0.015 | ||
1.25 μL | ±2.5 | ±0.031 | ±1.5 | ±0.019 | ||
2.5 μL | ±1.4 | ±0.035 | ±0.7 | ±0.018 | ||
| 0.5 μL | ±8.0 | ±0.04 | ±5.0 | ±0.025 | |
1 μL | ±2.5 | ±0.025 | ±1.8 | ±0.018 | ||
5 μL | ±1.5 | ±0.075 | ±0.8 | ±0.04 | ||
10 μL | ±1.0 | ±0.10 | ±0.4 | ±0.04 | ||
| 2 μL | ±5.0 | ±0.10 | ±1.5 | ±0.03 | |
10 μL | ±1.2 | ±0.12 | ±0.6 | ±0.06 | ||
20 μL | ±1.0 | ±0.20 | ±0.3 | ±0.06 | ||
| 2 μL | ±5.0 | ±0.10 | ±1.5 | ±0.03 | |
10 μL | ±1.2 | ±0.12 | ±0.6 | ±0.06 | ||
20 μL | ±1.0 | ±0.2 | ±0.3 | ±0.06 | ||
| 10 μL | ±3.0 | ±0.3 | ±0.7 | ±0.07 | |
50 μL | ±1.0 | ±0.5 | ±0.3 | ±0.15 | ||
100 μL | ±0.8 | ±0.8 | ±0.2 | ±0.2 | ||
| 20 μL | ±2.5 | ±0.5 | ±0.7 | ±0.14 | |
100 μL | ±1.0 | ±1.0 | ±0.3 | ±0.3 | ||
200 μL | ±0.6 | ±1.2 | ±0.2 | ±0.4 | ||
| 30 μL | ±2.5 | ±0.75 | ±0.7 | ±0.21 | |
150 μL | ±1.0 | ±1.5 | ±0.3 | ±0.45 | ||
300 μL | ±0.6 | ±1.8 | ±0.2 | ±0.6 | ||
| 100 μL | ±3.0 | ±3.0 | ±0.6 | ±0.6 | |
500 μL | ±1.0 | ±5.0 | ±0.2 | ±1.0 | ||
1000 μL | ±0.6 | ±6.0 | ±0.2 | ±2.0 | ||
| 0.25 mL | ±4.8 | ±0.012 | ±1.2 | ±0.003 | |
1.25 mL | ±0.8 | ±0.010 | ±0.2 | ±0.0025 | ||
2.5 mL | ±0.6 | ±0.015 | ±0.2 | ±0.005 | ||
| 0.5 mL | ±2.4 | ±0.012 | ±0.6 | ±0.003 | |
2.5 mL | ±1.2 | ±0.030 | ±0.25 | ±0.006 | ||
5.0 mL | ±0.6 | ±0.030 | ±0.15 | ±0.0075 | ||
| 1.0 mL | ±3.0 | ±0.030 | ±0.6 | ±0.006 | |
5.0 mL | ±0.8 | ±0.040 | ±0.2 | ±0.010 | ||
10.0 mL | ±0.6 | ±0.060 | ±0.15 | ±0.015 |